EasyManua.ls Logo

Rohde & Schwarz SMBV100B - Page 772

Rohde & Schwarz SMBV100B
1030 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SMBV100B
770User Manual 1178.4460.02 ─ 03
Manual operation: See "Carrier State" on page 236
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STARt <Start>
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:STOP <Stop>
Selects the last carrier in the carrier range to which the settings shall apply.
Parameters:
<Stop> integer
Range: 0 to 511
*RST: 0
Example:
BB:ARB:MCAR:EDIT:CARR:STOP 4
The carrier range stops at carrier 4.
Manual operation: See "Carrier Start/Start" on page 236
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
<Step>
Sets the step width by which the starting power of the carriers in the defined carrier
range is incremented.
Parameters:
<Step> float
Range: -80 to 80
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -80dB
Sets a power of -80 dB for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:POW:STEP 1 dB
The power is incremented by 1dB for each carrier.
That is, the first carrier has -80 dB, the second -79 dB, and so
on.
Manual operation: See "Gain Step" on page 236
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer[:STARt] <Start>
Sets the power for the individual carriers in the defined carrier range.
Parameters:
<Start> float
Range: -80 to 0
Increment: 0.01
*RST: 0
Default unit: dB
Example:
BB:ARB:MCAR:EDIT:CARR:POW -50 dB
sets the power of the carriers in the carrier range to -50 dB.
SOURce Subsystem

Table of Contents

Related product manuals