Configuring the RF Signal
R&S
®
SMBV100B
337User Manual 1178.4460.02 ─ 03
Remote command:
[:SOURce<hw>]:LIST:INDex on page 855
Mode
Selects the mode for list processing.
For detailed information on the sweep modes and the triggering, see Chapter 6.10.1,
"Signal Generation and Triggering in the Sweep and List Modes", on page 312.
"Auto"
Generates the signal by processing the frequency/level value pairs of
the list from the beginning to the end.
The list steps are performed automatically, controlled by the dwell
time. If you switch from any mode to "Auto", signal generation always
starts at the top of the list.
"Single / Extern Single"
Generates the signal by processing the frequency/level value pairs of
the list once from the beginning to the end after a trigger event.
The list steps are performed automatically, controlled by the dwell
time. If one cycle is completed, the instrument waits for the next trig-
ger event.
"Step / Extern Step"
Generates the signal by processing the frequency/level value pairs of
the list step-by-step, manually triggered.
Remote command:
[:SOURce<hw>]:LIST:TRIGger:SOURce on page 857
[:SOURce<hw>]:LIST:MODE on page 856
Dwell Time Mode
Selects either variable dwell times or a fixed dwell time to perform signal generation in
list mode.
"Global"
Uses the same dwell time for all lists steps, set with Global Dwell
Time.
"From List"
Uses the dwell times from the list.
You can define the dwell time for each frequency/level value pair indi-
vidually, see Chapter 6.10.7, "List Editor", on page 341.
Remote command:
[:SOURce<hw>]:LIST:DWELl:MODE on page 854
Global Dwell Time
Sets the dwell time for Dwell Time Mode > "Global".
See also "Dwell time" on page 325.
Remote command:
[:SOURce<hw>]:LIST:DWELl on page 853
Trigger Slope
For "Mode = Extern Step/Single", selects the polarity of the active slope of an applied
instrument trigger.
Trigger signal is expected at the Inst Trig connector.