Remote Control Commands
R&S
®
SMBV100B
670User Manual 1178.4460.02 ─ 03
Example: See Example "Connecting and configuring external instrument"
on page 662.
Manual operation: See "RF Couple" on page 448
:SCONfiguration:EXTernal:DIGital:RF:FREQuency <Frequency>
:SCONfiguration:EXTernal:IQOutput:RF:FREQuency <Frequency>
In uncoupled mode, sets the RF frequency of the external instrument.
Parameters:
<Frequency> float
Range: 100E3 to 3E9
Increment: 0.01
*RST: 1E9
Example:
SCONfiguration:EXTernal:IQOutput:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput:RF:FREQuency 2143E6
// RF frequency of the external instrument is RF = 2.143 GHz
Manual operation: See "(Delta) RF Frequency/ RF Level" on page 448
:SCONfiguration:EXTernal:DIGital:RF:FREQuency:OFFSet <FreqOffset>
:SCONfiguration:EXTernal:IQOutput:RF:FREQuency:OFFSet <FreqOffset>
In coupled mode, offsets the RF frequency of the external instrument with the selected
delta value.
Parameters:
<FreqOffset> float
Range: -3E9 to 3E9
Increment: 0.01
*RST: 0
Example:
SOURce1:FREQuency:CW 2143E6
SCONfiguration:EXTernal:IQOutput:RF:COUPling ON
SCONfiguration:EXTernal:IQOutput:RF:FREQuency:OFFSet 20E6
// The resulting RF frequency of the external instrument is
// RF = 2143E6 + 20E6 = 2.163 GHz
// Where both the RF frequency and the frequency offset
// are applied at the external instrument
Manual operation: See "(Delta) RF Frequency/ RF Level" on page 448
:SCONfiguration:EXTernal:DIGital:RF:POWer <Power>
:SCONfiguration:EXTernal:IQOutput:RF:POWer <Power>
In uncoupled mode, sets the RF level of the external instrument.
SCONfiguration Subsystem