Remote Control Commands
R&S
®
SMBV100B
671User Manual 1178.4460.02 ─ 03
Parameters:
<Power> float
Range: -130 to 20
Increment: 0.01
*RST: -30
Example:
SOURce1:POWer:LEVel:IMMediate:AMPLitude -30
SCONfiguration:EXTernal:IQOutput:RF:COUPling OFF
SCONfiguration:EXTernal:IQOutput:RF:POWer -20
// RF level of the external instrument is -20 dB
Manual operation: See "(Delta) RF Frequency/ RF Level" on page 448
:SCONfiguration:EXTernal:DIGital:RF:POWer:OFFSet <PowerOffset>
:SCONfiguration:EXTernal:IQOutput:RF:POWer:OFFSet <PowerOffset>
In coupled mode, offsets the RF level of the external instrument with the selected delta
value.
Parameters:
<PowerOffset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Example:
SOURce1:POWer:LEVel:IMMediate:AMPLitude -30
SCONfiguration:EXTernal:IQOutput:RF:COUPling ON
SCONfiguration:EXTernal:IQOutput:RF:POWer:OFFSet -20
// RF level of the external instrument is -50 dB
// Where both the RF level and the level offset
// are applied at the external instrument
Manual operation: See "(Delta) RF Frequency/ RF Level" on page 448
:SCONfiguration:EXTernal:DIGital:RF:STATe <RemConnState>
:SCONfiguration:EXTernal:IQOutput:RF:STATe <RemConnState>
Queries/sets the RF output state of the connected external instrument.
Parameters:
<RemConnState> 0 | 1 | OFF | ON
Example:
SCONfiguration:EXTernal:IQOutput:RF:COUPling ON
OUTput1:STATe ON
SCONfiguration:EXTernal:IQOutput:RF:STATe?
// 1
Example:
SCONfiguration:EXTernal:DIGital:RF:COUPling OFF
SCONfiguration:EXTernal:DIGital:RF:STATe ON
Manual operation: See "RF State" on page 449
SCONfiguration Subsystem