Remote Control Commands
R&S
®
SMBV100B
679User Manual 1178.4460.02 ─ 03
:OUTPut:USER<ch>:SIGNal <Signal>
Sets the control signal that is output at the selected connector.
To define the connector direction, use the command :OUTPut:USER<ch>:
DIRection.
Suffix: 
USER<ch>
.
1 to 5
Parameters:
<Signal> MARKA1 | MARKA2 | MARKA3 | NONE | SYNCOUT | SCLock |
LATTenuation | BGATe | HOP | CWMODulation | TRIGgered |
MTRigger | LOW | HIGH | BERRESTOUT | BERDATENOUT |
BERCLKOUT | BERDATOUT
MARKA<1|2|3> = Baseband Marker 1/2/3
MTRigger = Manual Trigger, available for USER5
BERRESTOUT|BERDATENOUT|BERCLKOUT|BERDATOUT = BER
TestGen Data, Clock, Data Enable and Restart
SYNCOUT = Baseband Sync Out
SCLock = Symbol Clock
LATTenuation = Lev Att
BGATA = Burst Gate
HOP = HOP
CWMODulation = CW/Mod
TRIGgered = Triggered
MTRigger = Manual Trigger, available for USER5
LOW|HIGH = Always 0/1
NONE = none
*RST: 
MARKA1
Example:  See Example "Global connectors settings configuration"
on page 677
Options:  BERRESTOUT|BERDATENOUT|BERCLKOUT|BERDATOUT
require R&S SMBVB-K80
Manual operation:  See "Signal" on page 522
:OUTPut:USER<ch>:TRIGger[:IMMediate]
Generates a short pulse signal and outputs it at the User connector.
This signal can serve as a common external trigger signal for triggering of several
R&S SMBV100B, see Example "Triggering several R&S SMBV100B instruments
simultaneously" on page 434.
Suffix: 
USER<ch>
.
5
Example:  See Example "Global connectors settings configuration"
on page 677.