Remote Control Commands
R&S
®
SMBV100B
706User Manual 1178.4460.02 ─ 03
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate
If [:SOURce<hw>]:PULM:TRIGger:MODE SINGle, triggers the pulse generator.
Example:  See Example "Perform pulse modulation" on page 704.
Manual operation:  See "Execute Single Trigger" on page 362
[:SOURce<hw>]:PULM:PERiod <Period>
Sets the period of the generated pulse, that means the repetition frequency of the inter-
nally generated modulation signal.
Parameters:
<Period> float
The minimum value depends on the installed options
R&S SMBVB-K22 or R&S SMBVB-K23
Range:  20E-9  to  100
Increment:  5E-9
*RST:  10E-6
Example:  See Example "Perform pulse modulation" on page 704.
Manual operation:  See "Pulse Period" on page 361
[:SOURce<hw>]:PULM:WIDTh <Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width> float
Range:  20E-9  to  100
Increment:  10E-9
*RST:  2E-6
Example:  See Example "Perform pulse modulation" on page 704.
Manual operation:  See "Pulse Width" on page 361
[:SOURce<hw>]:PULM:DELay <Delay>
Sets the pulse delay.
Parameters:
<Delay> float
*RST:  1ms
Example:  See Example "Perform pulse modulation" on page 704.
Manual operation:  See "Pulse Delay" on page 362
SOURce Subsystem