[:SOURce<hw>]:PULM:DOUBle:DELay <Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay> float
*RST: 1E-6
Example: See Example "Perform pulse modulation" on page 704.
Manual operation: See "Double Pulse Delay" on page 362
[:SOURce<hw>]:PULM:DOUBle:WIDTh
<Width>
Sets the width of the second pulse.
Parameters:
<Width> float
Increment: 5E-9
Example: See Example "Perform pulse modulation" on page 704.
Manual operation: See "Double Pulse Width" on page 362
[:SOURce<hw>]:PULM:DOUBle:STATe <State>
Provided for backward compatibility with former Rohde & Schwarz signal generators.
Works like the command [:SOURce<hw>]:PULM:MODE DOUBle.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
[:SOURce<hw>]:PULM:STATe <State>
Activates pulse modulation.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Perform pulse modulation" on page 704.
Manual operation: See "State" on page 350
[:SOURce<hw>]:PULM:SOURce <Source>
Selects between the internal (pulse generator) or an external pulse signal for the mod-
ulation.
SOURce Subsystem