Remote Control Commands
R&S
®
SMBV100B
763User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:BB:ARBitrary:MCARrier:SETTing:STORe <Filename>
Stores the current settings into the selected file; the file extension (*.arb_multcarr)
is assigned automatically.
Setting parameters:
<Filename> string
Filename or complete file path
Example: See Example "Generating a multi carrier signal" on page 744
Usage: Setting only
Manual operation: See "Save/Recall Frame" on page 228
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:COUNt
<Count>
Sets the number of carriers in the ARB multi-carrier waveform.
Parameters:
<Count> integer
Range: 1 to 512
*RST: 1
Example: See Example "Generating a multi carrier signal" on page 744
Manual operation: See "Number of Carriers" on page 228
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:MODE <Mode>
The command sets the carrier frequency mode.
Parameters:
<Mode> EQUidistant | ARBitrary
*RST: EQUidistant
Example:
BB:ARB:MCAR:CARR:MODE EQU
Sets an equidistant carrier spacing. The carrier frequency can-
not be set.
Manual operation: See "Mode" on page 228
[:SOURce<hw>]:BB:ARBitrary:MCARrier:CARRier:SPACing <Spacing>
Sets the frequency spacing between adjacent carriers of the multi-carrier waveform
(see Chapter 3.8.2.1, "Defining the Carrier Frequency", on page 225).
SOURce Subsystem

Table of Contents