Remote Control Commands
R&S
®
SMBV100B
902User Manual 1178.4460.02 ─ 03
:SYSTem:COMMunicate:SERial:BAUD <Baud>
Defines the baudrate for the serial remote control interface.
Parameters:
<Baud> 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200
*RST: n.a. (factory preset: 115200)
Example: See Example "Finding out the used VISA ressource strings"
on page 890.
Manual operation: See "Baud Rate" on page 566
:SYSTem:COMMunicate:SERial:PARity <Parity>
Enters the parity for the serial remote control interface.
Parameters:
<Parity> NONE | ODD | EVEN
*RST: n.a. (factory preset: NONE)
Example: See Example "Finding out the used VISA ressource strings"
on page 890.
Manual operation: See "Parity" on page 566
:SYSTem:COMMunicate:SERial:RESource?
Queries the visa resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource> string
Example: See Example "Finding out the used VISA ressource strings"
on page 890.
Usage: Query only
:SYSTem:COMMunicate:SERial:SBITs <SBits>
Defines the number of stop bits for the serial remote control interface.
Parameters:
<SBits> 1 | 2
*RST: n.a. (factory preset: 1)
Example: See Example "Finding out the used VISA ressource strings"
on page 890.
Manual operation: See "Stop Bits" on page 567
SYSTem Subsystem