Remote Control Commands
R&S
®
SMBV100B
919User Manual 1178.4460.02 ─ 03
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:SOURce AUTO
// detects the connectors at that signal is assigned
// and performs the connecton test
Manual operation: See "Source" on page 936
:TEST:BB:BNC:DESTination <BncDestination>
Selects the BNC connection test destination.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Parameters:
<BncDestination> AUTO | USER1 | USER2 | USER3 | USER4 | USER5
*RST: USER2
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:DESTination USER1
Manual operation: See "Destination" on page 936
:TEST:BB:BNC:LOG?
Queries the log message reported during the BNC connector test.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
Return values:
<Log> string
Example:
SYSTem:PROTect1 0,123456
TEST:BB:BNC:LOG?
// queries the reported message
Usage: Query only
Manual operation: See "Log Message" on page 937
:TEST:BB:BNC:SOURce <BncSource>
Selects the BNC connection test source.
This is a password-protected function. Unlock the protection level 1 to access it, see :
SYSTem:PROTect<ch>[:STATe].
TEST Subsystem