EasyManua.ls Logo

Inficon HAPSITE Smart - Page 90

Inficon HAPSITE Smart
554 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 - 2
IPN 074-397-P1G
HAPSITE Smart Operating Manual
Figure 3-1 The Power Switch
3 To warm up the system and perform a tune check, a method must be chosen
to run, unless a startup method has been activated. Press 1, then SEL. See
Figure 3-2.
Figure 3-2 Main Menu
HINT: The HAPSITE can be configured to use a startup method. If a startup
method is chosen, Step 3 on page 3-2 through Step 5 on page 3-5 will
be automatically run. Currently, instruments are shipped with the
Survey method selected as the startup method.
4 Select a method to run.
To select a menu option:
use up and down arrows to highlight the desired method, press SEL.
OR
press the number of the selection and then press SEL.
HINT: Refer to Section 3.3, Survey Mode, on page 3-18 or Section 3.4,
GC/MS Mode with Loop, on page 3-23 for additional instructions.
HINT: This example will use a GCMS Method with the sample loop installed.
See Figure 3-3.

Table of Contents

Related product manuals