103
Chapter 5
5.3 Exposure with DR Detector (Aero Sync mode)
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When the "Ready" is displayed, immediately perform
exposure.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If image acquisition started without performing ex-
posure, this may result in incorrect detection due to
noise, etc. Perform rejecting (image deletion) and re-
peat exposure.
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
z
When the [AeroSync Button] is ON
1
Prepare the DR Detector.
A message "Waiting for permission" is displayed
in the image display area.
HINT
•••••••••••••••••••••••••••••••••••••
If the information about the body part is not imported
while importing examination information from the high-
er system, select the body part to be imaged. Then
press [Perform] on the body part selection screen to
display the exposure screen.
When the AeroDR power management level is set to
[3], [4] or [5], "Waiting to start shooting" is displayed
instead of "Waiting for permission".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
IMPORTANT
•••••••••••••••••••••••••••••••••••••
A malfunction may be caused by a shock to the DR
Detector. In such cases, set [AeroSync Button] to ON
on the system monitor screen.
To prevent incorrect operations, always check that
"Ready" is displayed.
Conditions for changing the detection level vary de-
pending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
The X-ray detection level can be set when making an
exposure. Refer to "5.3 Exposure with DR Detector
(Aero Sync mode)" for information on the X-ray detec-
tion levels that can be set.
When importing examination information from the
higher system and when exposure body part informa-
tion is not imported, select the exposure body part.
When pressing [Perform] on the body part selection
screen, the exposure screen is displayed.
When you move the focus in the image display area
on the exposure resumption waiting screen to cap-
ture an image of another body part, not the ready-to-
expose screen but the exposure resumption waiting
screen is displayed.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Perform exposure.
Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch
As soon as the exposed image is transferred,
the image is displayed in the image display area.