104
5.3 Exposure with DR Detector (Aero Sync mode)
2
After positioning the patient, check the X-
ray detection level, then press [AeroSync].
• If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
• Change the X-ray detection level as needed.
• After checking the X-ray detection level, press
[AeroSync].
• A message "Ready" is displayed in the mes-
sage/button area. The screen is locked.
• The screen is locked. To return to the previous
screen to change the position of the patient,
etc., press [Cancel].
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• To prevent incorrect operations, always press
[AeroSync] and check that "Ready" is displayed.
• Conditions for changing the detection level vary
depending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
• When the AeroDR power management level is set to
[3], [4] or [5], it may take some time until the exposure
becomes ready after you have pressed [Aero Sync]. In
this case, [Standby] button instead of [Cancel] button
is displayed in the message/button area.
• After pressing [AeroSync], do not touch the wired
connector of the DR Detector until the image transfer
process completes.
• When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
• The X-ray detection level can be set when making an
exposure. Refer to "5.3 Exposure with DR Detector
(Aero Sync mode)" for information on the X-ray detec-
tion levels that can be set.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Perform exposure.
• Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch
•
As soon as the exposed image is transferred,
the image is displayed in the image display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When the "Ready" is displayed, immediately perform
exposure.
• Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
• If image acquisition started without performing ex-
posure, this may result in incorrect detection due to
noise, etc. Perform rejecting (image deletion) and re-
peat exposure.
• If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
• When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
• When linked with the X-ray device, because the se-
lected exposure conditions at the time of X-ray expo-
sure are linked to the irradiation results, always check
the exposure conditions before X-ray exposure.
• Depending on the type of X-ray device, several sec-
onds may elapse before notication of the irradiation
results. For this type of model, take care because
sometimes irradiation results are not reected correct-
ly if the exposure conditions are changed immediately
after X-ray irradiation.
•••••••••••••••••••••••••••••••••••••••••••••••••••••