104
5.3 Exposure with DR Detector (Aero Sync mode)
2
After positioning the patient, check the X-
ray detection level, then press [AeroSync].
If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
Change the X-ray detection level as needed.
After checking the X-ray detection level, press
[AeroSync].
A message "Ready" is displayed in the mes-
sage/button area. The screen is locked.
The screen is locked. To return to the previous
screen to change the position of the patient,
etc., press [Cancel].
IMPORTANT
•••••••••••••••••••••••••••••••••••••
To prevent incorrect operations, always press
[AeroSync] and check that "Ready" is displayed.
Conditions for changing the detection level vary
depending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
When the AeroDR power management level is set to
[3], [4] or [5], it may take some time until the exposure
becomes ready after you have pressed [Aero Sync]. In
this case, [Standby] button instead of [Cancel] button
is displayed in the message/button area.
After pressing [AeroSync], do not touch the wired
connector of the DR Detector until the image transfer
process completes.
When using the AeroDR 3 1417HD, click the modal-
ity icon before making an exposure and select the
AeroDR 3 1417HD.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
The X-ray detection level can be set when making an
exposure. Refer to "5.3 Exposure with DR Detector
(Aero Sync mode)" for information on the X-ray detec-
tion levels that can be set.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
3
Perform exposure.
Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch
As soon as the exposed image is transferred,
the image is displayed in the image display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When the "Ready" is displayed, immediately perform
exposure.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If image acquisition started without performing ex-
posure, this may result in incorrect detection due to
noise, etc. Perform rejecting (image deletion) and re-
peat exposure.
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
When linked with the X-ray device, because the se-
lected exposure conditions at the time of X-ray expo-
sure are linked to the irradiation results, always check
the exposure conditions before X-ray exposure.
Depending on the type of X-ray device, several sec-
onds may elapse before notication of the irradiation
results. For this type of model, take care because
sometimes irradiation results are not reected correct-
ly if the exposure conditions are changed immediately
after X-ray irradiation.
•••••••••••••••••••••••••••••••••••••••••••••••••••••