7.3 Exposure with AeroDR Portable UF Unit
Chapter 7
IMPORTANT
•••••••••••••••••••••••••••••••••••••
To prevent incorrect operations, always press
[AeroSync] and check that "Ready" is displayed.
Conditions for changing the detection level vary
depending on the X-ray device being used. If a survey
has been carried out when the system was installed,
follow the survey results.
When the AeroDR power management level is set to
[3], [4] or [5], it may take some time until the exposure
becomes ready after you have pressed [Aero Sync]. In
this case, [Standby] button instead of [Cancel] button
is displayed in the message/button area.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••
Refer to "7.3 Exposure with AeroDR Portable UF Unit"
for information on the X-ray detection levels that can
be set for exposures.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
5
Perform exposure.
Congure the portable X-ray device settings and
check the exposure setting before making an
exposure.
Press the exposure switch
As soon as the exposed image is transferred,
the image is displayed in the image display area.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When performing an exposure, stay away from the patient
as far as possible before pressing the exposure switch.
When the "Ready" is displayed, immediately perform
exposure.
Do not let the DR Detector vibrate or receive shock
until the exposed image is displayed.
If image acquisition started without performing exposure,
this may result in incorrect detection due to noise, etc.
Perform reject (image deletion) and repeat exposure.
If image acquisition does not start after exposure, the
voltage may not be enough. Check the exposure con-
ditions and repeat exposure.
When exposure is done while "Ready" is not dis-
played, have an interval of 1 minute or more before
repeating exposure.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
6
Adjust the image.
After exposure, adjust the image as required.
7
Press [ ].
The exposure is completed, and the display re-
turns to the initial screen.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
When consolidating image adjustment, press [
] to
pause the examination and return to the initial screen.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
8
When exposure is nished, turn the Aero-
DR Portable UF Unit o󰀨.
Turn off the power switch of the AeroDR Por-
table UF Unit and confirm that the power LED
(green/orange) is OFF
.
For shutdown operation of the AeroDR Portable
UF Unit, refer to the "AeroDR Portable UF Unit
Operation Manual".
HINT
•••••••••••••••••••••••••••••••••••••
Image adjustment can be done on the portable de-
vice's [Image Proc] tab. However, if a more advanced
image adjustment is required, perform image adjust-
ment on the [Image Proc] tab or the viewer screen.
For details on the viewer screen, refer to "6.7 Func-
tions of the viewer screen".
When moving the exposure location, turn off the
AeroDR Portable UF Unit. After moving the exposure
location, perform the procedure again from step 1.
It is all right to keep power of this device and the DR
Detector on.
•••••••••••••••••••••••••••••••••••••••••••••••••••••