7.3 Exposure with AeroDR Portable UF Unit
z
When the [AeroSync Button] is ON
1
When you arrive at the exposure location,
turn on the power of the AeroDR Portable
UF Unit.
• Turn on the power switch of the AeroDR Por-
table UF Unit and conrm that the (green) power
LED is ON.
• For startup operation of the AeroDR Portable UF
Unit, refer to "AeroDR Portable UF Unit Opera-
tion Manual".
• Conrm that the icon of the DR Detector to be
used in exposure is displayed on the application
bar.
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When the AP information channel is set to [Auto CH],
press [Auto CH] on the system monitor screen (por-
table) and check the channel.
• It may take several minutes for wireless connection to
start after turning on the AeroDR Portable UF Unit.
•••••••••••••••••••••••••••••••••••••••••••••••••••••
2
Select the examination.
3
Press [Perform].
• The exposure screen is displayed.
• A message "Waiting for permission" is displayed
in the image display area.
HINT
•••••••••••••••••••••••••••••••••••••
• If the information about the body part is not imported
while importing examination information from the high-
er system, select the body part to be imaged. Then
press [Perform] on the body part selection screen to
display the exposure screen.
• When the AeroDR power management level is set to
[3], [4] or [5], "Waiting to start shooting" is displayed
instead of "Waiting for permission".
•••••••••••••••••••••••••••••••••••••••••••••••••••••
4
After positioning the patient, check the X-
ray detection level, then press [AeroSync].
• If the control panel does not switch to
[AeroSync], press the [AeroSync] tab.
• Change the X-ray detection level as needed.
• After checking the X-ray detection level, press
[AeroSync].
• A message "Ready" is displayed in the mes-
sage/button area. The screen is locked.
• The screen is locked. To return to the previous
screen to change the position of the patient,
etc., press [Cancel].