10.2 Error codes and countermeasures
Error code Displayed message Problems/countermeasures
C36022
At PrimtComposer process, processing is timeout.
(1) Try performing the operation again.
(2) Contact Konica Minolta technical representatives.
C37001 [DR Detector name] Connecting detector
Check the network equipment and communication
cable connection status.
C37002 [DR Detector name] disconnection error
C37003
C37004
Error. Please touch the work area.
Check the network equipment and communication
cable connection status.
Touching either the working area or exposure condi-
tion items may x the problem.
C37005
[DR Detector name] Disconnected energy saving not
set. Please touch the work area.
(1) Touch either the working area or exposure condi-
tion items.
(2) Check the network equipment and communication
cable connection status.
(3) Stop the examination.
C37006 [DR Detector name] wireless forbidden error
Check the network equipment and communication
cable connection status.
C37007
[DR Detector name] version up error
Version upgrade failed.
Check the connection and restart the AeroDR cas-
sette.
C37008
[DR Detector name] calibration error
Calibration failed.
Check the connection and retry.
(1) Check the network equipment and communication
cable connection status.
(2) Restart the calibration tool.
C37009
There are images unreceived to the AeroDR.
Do not turn o󰀨 the power of AeroDR.
Return AeroDR detector to previous connection
state(Wired/Wireless). In the case of wireless connec-
tion, the electric wave state has deteriorated.
Please check an electric wave state by the AeroDR
icon at the upper right of a screen, and wait until con-
nection returns.
(1) Check the network equipment and communication
cable connection status.
(2) Press [OK], and wait until image re-acquisition is
complete.
C37010
[DR Detector name] Correction processing error
Correction processing error.
Restart the CS-7 and retake.
Restart this device and perform the exposure again.
C37012 [DR Detector name] Additional registration error
Check the network equipment and communication
cable connection status.
C37013 [DR Detector name] Registration cancelation error
C37014 [DR Detector name] Set panel data error
C37015 [DR Detector name] Get panel data error
C37017 [DR Detector name] Attribute information error.
C37018
[DR Detector name] Error in image le creation.
Please contact the service representation.
Contact Konica Minolta technical representatives.
C37019
[DR Detector name] Exposure Permission Setting Er-
ror
Please touch the work area.
(1) Touch either the working area or exposure condi-
tion items.
(2) Check the network equipment and communication
cable connection status.
C37025
[DR Detector name] Exposure Mode Setting Error
Please touch the work area.
Touching either the working area or exposure condi-
tion items may x the problem. If not, contact Konica
Minolta technical representatives.
C37054
Exposure eld setting error.
Check the network connection and touch this expo-
sure condition.
(1) Check the network equipment and communication
cable connection status.
(2) When [Ready] is displayed, press the Exam. Tag.
C37055
Communication error.
Check the network connection and touch this expo-
sure condition.
C37056
Version up failed.
Check the connection and restart the AeroDR XG BOX/
AeroDR Interface Unit2.
(1) Check the network equipment and communication
cable connection status.
(2) Restart the device and the AeroDR Interface Unit.
C37058
Failed to set the Setting data.
Please check network connection and retry.
Contact Konica Minolta technical representatives.
C37059
Failed to get the Setting data.
Please check network connection and retry.
Contact Konica Minolta technical representatives.