398
10.2 Error codes and countermeasures
Error code Displayed message Problems/countermeasures
S008**
S00900
S00910
S00920
S00930
[AeroDR Stitching Unit name] MCB Error
Contact a service representative.
Contact Konica Minolta technical representatives.
S00940
S00A00
S00A10
S00A20
[AeroDR Stitching Unit name] Motor Error
Please power OFF/ON of AeroDR Stitching unit.
Restart the AeroDR Stitching System.
S00A30
[AeroDR Stitching Unit name] Sensor Error
Please power OFF/ON of AeroDR Stitching unit.
S00A40
S00A50
[AeroDR Stitching Unit name] Motor Error
Please power OFF/ON of AeroDR Stitching unit.
S00A60
S00C90
[AeroDR Stitching Unit name] Sensor Error
Please power OFF/ON of AeroDR Stitching unit.
S00CA0
[AeroDR Stitching Unit name] Side Cover Open
Please close the AeroDR stitching unit cover and
press RESET switch.
Close the side cover of the AeroDR Stitching Unit, and
press the Reset switch on the AeroDR Stitching X-Ray
Auto-barrier Unit to cancel the error.
S00CB0
[AeroDR Stitching Unit name] Panel Error
Properly set the grid inside AeroDR stitching unit,
close the over and press the RESET switch.
Check that the DR Detector is correctly inserted in the
AeroDR Stitching Unit, and close the side cover. Press
the Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error.
S00CC0
[AeroDR Stitching Unit name] Grid Error
Properly set the grid inside AeroDR stitching unit,
close the over and press the RESET switch.
Check that the grid is correctly installed on the AeroDR
Stitching Unit, and then close the side cover. Press
the Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error.
S00CD0
[AeroDR Stitching Unit name] Panel Error
Properly close the inner cover of AeroDR stitching unit
and switch the power OFF/ON.
(1) Open the side cover on the AeroDR Stitching Unit
and close the internal lid.
(2) Close the side cover and restart the AeroDR Stitch-
ing System.
S200**
[AeroDR Stitching Unit name] Internal connection error
Please power OFF of AeroDR Stitching unit.
Please confirm the connectivity of Power box and auto
barrier unit before switching the power to ON.
(1) Turn o the AeroDR Stitching System.
(2) Check the power unit and the AeroDR Stitching X-
Ray Auto-barrier Unit are connected by the power
cable, and turn on the AeroDR Stitching System.
S208**
[AeroDR Stitching Unit name] Internal Error
Please power OFF/ON of AeroDR Stitching unit.
Please confirm the connectivity of Power box and
Stitiching unit before switching the power to ON.
(1) Turn o the AeroDR Stitching System.
(2) Check the power unit and the AeroDR Stitching
Unit are connected by the power cable, and turn
on the AeroDR Stitching System.
S20C50
[AeroDR Stitching Unit name] Network Error
Please confirm the network of AeroDR stitching unit
and CS-7, switch on the power OFF/ON of AeroDR
stitching unit.
Check that the AeroDR Stitching Unit and this device
are connected by the network cable and restart the
AeroDR Stitching System.
S20C60
[AeroDR Stitching Unit name] Network Error
Please confirm the network of AeroDR stitching unit
and CS-7, press the AeroDR Stitching unit RESET
switch.
Check that the AeroDR Stitching Unit and this device
are connected by the network cable, and press the
Reset switch on the AeroDR Stitching X-Ray Auto-
barrier Unit to cancel the error.
S30460
[AeroDR Stitching Unit name] Height detection (out of
range)
Please press the AeroDR Stitching unit RESET switch.
Repeat the height detection.
(1) Press the Reset switch on the AeroDR Stitching
X-Ray Auto-barrier Unit to cancel the error.
(2) Check the collimator installation position in the X-
ray device is correct.
(3) Execute height detection.
(4) If the problem is not solved by the above method,
there is a possibility that the intensity of the expo-
sure eld lamp has changed. Change the bright-