92
5.2 Exposure procedure with DR Detector (Basic connection)
4
Take the 2nd image.
• Configure the X-ray device settings and check
the exposure setting before making an expo-
sure.
Press the exposure switch
(1)
(3)
(2)
•
The exposed image is displayed in image display
area (2), and the focus moves to (3). At this stage,
because the DR Detector is inserted in the wall
stand, exposure of (3) is not possible.
5
Remove the DR Detector from the wall
stand, and prepare for exposure with a
stand-alone DR Detector.
(1)
(3)
(2)
• When the DR Detector is removed from the wall
stand, "Preparing" is displayed in image display
area (3).
6
Take 3rd image.
• Press the exposure switch after "Ready" is dis-
played in image display area (3).
•
Congure the X-ray device settings and check the
exposure setting before making an exposure.
Press the exposure switch
(1)
(3)
(2)
(1)
(3)
(2)
•
The exposed image is displayed in image display
area (3). The focus remains on image display area (3).
IMPORTANT
•••••••••••••••••••••••••••••••••••••
• When linked with the X-ray device, because the se-
lected exposure conditions at the time of X-ray expo-
sure are linked to the irradiation results, always check
the exposure conditions before X-ray exposure.
• Depending on the type of X-ray device, several sec-
onds may elapse before notication of the irradiation
results. For this type of model, take care because
sometimes irradiation results are not reected correct-
ly if the exposure conditions are changed immediately
after X-ray irradiation.
•••••••••••••••••••••••••••••••••••••••••••••••••••••