Remote Control Commands
R&S
®
SMBV100B
826User Manual 1178.4460.02 ─ 03
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01Hz
*RST: 100 MHz
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 882
Manual operation: See " Current Frequency" on page 328
[:SOURce<hw>]:FREQuency:MULTiplier <Multiplier>
Sets the multiplication factor N
FREQ:MULT
of a subsequent downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
set with the command FREQ.
The query FREQ? returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See "RF frequency and level display with a downstream instrument" on page 288.
Parameters:
<Multiplier> float
Range: 1 to dynamic
Increment: 0.001
*RST: 1
Example: See Example "Frequency configuration" on page 820
Manual operation: See "Multiplier" on page 291
[:SOURce<hw>]:FREQuency:OFFSet <Offset>
Sets the frequency offset f
FREQ:OFFSet
of a downstream instrument.
The parameters offset f
FREQ:OFFSer
and multiplier N
FREQ:MULT
affect the frequency value
set with the command FREQ.
The query FREQ? returns the value corresponding to the formula:
f
FREQ
= f
RFout
* N
FREQ:MULT
+ f
FREQ:OFFSer
See "RF frequency and level display with a downstream instrument" on page 288.
Note: The offset also affects RF frequency sweep.
Parameters:
<Offset> float
Increment: 0.01
*RST: 0
Example: See Example "Frequency configuration" on page 820
Manual operation: See "Offset" on page 291
SOURce Subsystem