EasyManua.ls Logo

Tektronix Keithley 2601B-PULSE - Configuring trigger attributes in local mode; Front-panel triggering example

Tektronix Keithley 2601B-PULSE
849 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 4: Sourcing and measuring Model 2601B-PULSE System SourceMeter Instrument Reference Manual
4-44 2601B-PULSE-901-01A April 2020
Configuring trigger attributes in local mode
From the front panel, press the CONFIG key and select TRIG. These menu items are available:
• TRIGGER-IN: Use these options to select the trigger-in source:
â–ª IMMEDIATE: Triggering occurs immediately and the instrument starts to make
measurements when it is ready (for example, after the source output is turned on).
â–ª MANUAL: The front-panel TRIG key must be pressed to trigger the instrument to
make readings.
• COUNT: Sets the trigger count (number of measurements) as follows:
â–ª FINITE: The instrument goes through measurement cycles for the programmed trigger count
(1 to 99999).
â–ª INFINITE: The instrument goes through measurement cycles indefinitely until halted.
• INTERVAL: Sets the time interval between measurements (0 s to 999.999 s) when the count is
greater than 1.
• DELAY: Sets the delay period between the trigger and the start of measurement
(0 s to 999.999 s).
Front-panel triggering example
This example uses the front panel to configure the trigger parameters to meet the following
requirements:
• Manual triggering (TRIG key)
• Infinite trigger count (cycle indefinitely through measurement cycles)
• Interval (time between measurements): 1 s
• Delay (time from trigger to measurement): 2 s
To configure the trigger parameters:
1. Press the CONFIG key, and then the TRIG key.
2. Select TRIGGER-IN, and then press the ENTER key or the navigation wheel .
3. Select MANUAL, and then press the ENTER key or the navigation wheel .
4. Select COUNT, then select INFINITE, and then press the ENTER key or the navigation wheel .
5. Select INTERVAL, set the interval to 1 s, and then press the ENTER key or the navigation
wheel .
6. Choose DELAY, set the delay to 2 s, and then press the ENTER key.
7. Use the EXIT (LOCAL) to return to the normal display.
8. Press the OUTPUT ON/OFF control to turn the output on.
9. Press TRIG. A 2 s delay occurs before the first measurement. The instrument cycles through
measurements indefinitely with a 1 s interval between measurements.
10. Press the OUTPUT ON/OFF control again to stop making readings.

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Related product manuals