Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-67
display.screen
This attribute contains the selected display screen.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Saved setup
0 (display.SMUA)
Usage
displayID = display.screen
display.screen = displayID
displayID
One of the following values:
â–ª 0 or display.SMUA: Displays source-measure and compliance values
â–ª 3 or display.USER: Displays the user screen
Details
Setting this attribute selects the display screen for the front panel. This performs the same action as
pressing the DISPLAY key on the front panel. The text for the display screen is set by
display.settext().
Read this attribute to determine which of the available display screens was last selected.
Example
display.screen = display.SMUA
Selects the source-measure and compliance
limit display.
Also see
display.settext() (on page 11-70)
display.sendkey()
This function sends a code that simulates the action of a front-panel control.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.sendkey(keyCode)
keyCode
A parameter that specifies the key press to simulate; see Details for
more information
Details
This command simulates pressing a front-panel key or navigation wheel, or turning the navigation
wheel one click to the left or right.