Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-67
display.screen
This attribute contains the selected display screen.
Instrument reset
Recall setup
Usage
displayID = display.screen
display.screen = displayID
One of the following values:
â–ª 0 or display.SMUA: Displays source-measure and compliance values
â–ª 3 or display.USER: Displays the user screen
Details
Setting this attribute selects the display screen for the front panel. This performs the same action as
pressing the DISPLAY key on the front panel. The text for the display screen is set by
display.settext().
Read this attribute to determine which of the available display screens was last selected.
Example
display.screen = display.SMUA
Selects the source-measure and compliance
limit display.
Also see
display.settext() (on page 11-70)
display.sendkey()
This function sends a code that simulates the action of a front-panel control.
Usage
display.sendkey(keyCode)
A parameter that specifies the key press to simulate; see Details for
more information
Details
This command simulates pressing a front-panel key or navigation wheel, or turning the navigation
wheel one click to the left or right.