Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-386 2601B-PULSE-901-01A April 2020
Details
This function sets and gets the termination character sequence that is used to indicate the end of a
line for a TSP-Net connection.
Using the termSequence parameter sets the termination sequence. The present termination
sequence is always returned.
For the termSequence parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is tspnet.TERM_LF. For devices that are not
TSP-enabled, the default is tspnet.TERM_CRLF.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID, tspnet.TERM_LF)
end
Sets termination type for IP address
192.0.2.1 to TERM_LF.
Also see
tspnet.connect() (on page 11-380)
tspnet.disconnect() (on page 11-381)
tspnet.timeout
This attribute sets the timeout value for the tspnet.connect(), tspnet.execute(), and tspnet.read()
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Instrument reset
Recall setup
Not saved
20.0 (20 s)
Usage
value = tspnet.timeout
tspnet.timeout = value
value
The timeout duration in seconds (1 ms to 30.0 s)
Details
This attribute sets the amount of time the tspnet.connect(), tspnet.execute(), and
tspnet.read() commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution but is only
accurate to the nearest 10 ms.
Example
tspnet.timeout = 2.0
Sets the timeout duration to 2 s.
Also see
tspnet.connect() (on page 11-380)
tspnet.execute() (on page 11-382)
tspnet.read() (on page 11-383)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals