Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-24 2601B-PULSE-901-01A April 2020
bufferVar.collecttimestamps
This attribute sets whether or not timestamp values are stored with the readings in the buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
See Details
0 (disabled)
Usage
state = bufferVar.collecttimestamps
bufferVar.collecttimestamps = state
state
Timestamp value collection status; set to one of the following:
â–ª 0: Timestamp value collection disabled (off)
â–ª 1: Timestamp value collection enabled (on)
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a dedicated
reading buffer
Details
Assigning a value to this attribute enables or disables the storage of timestamps. Reading this
attribute returns the state of timestamp collection.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.
When on, timestamp values are stored with readings in the buffer. This requires four extra bytes of
storage for each reading. Turning on additional collection items, such as timestamps (this attribute)
and source values, decreases the capacity of a dedicated reading buffer (for example,
smua.nvbuffer1), but does not change the capacity of a user-defined dynamically allocated buffer.
This value, off (0) or on (1), can only be changed when the buffer is empty. Empty the buffer using the
bufferVar.clear() function.
Example
smua.nvbuffer1.collecttimestamps = 1
Include timestamps with readings for
dedicated reading buffer 1.
Also see
bufferVar.clear() (on page 11-21)
Reading buffers (on page 5-1)
smua.measure.overlappedY() (on page 11-205)
smua.measure.Y() (on page 11-209)
smua.nvbufferY (on page 11-211)
smua.trigger.measure.Y() (on page 11-251)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals