Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-162 2601B-PULSE-901-01A April 2020
script.restore()
This function restores a script that was removed from the runtime environment.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
script.restore(name)
name
The name of the script to be restored
Details
This command copies the script from nonvolatile memory into the runtime environment. It also creates
a global variable with the same name as the name of the script.
Example
script.restore("test9")
Restores a script named test9 from
nonvolatile memory.
Also see
script.delete() (on page 11-158)
script.run()
This function runs the anonymous script.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
script.run()
run()
Details
Each time the script.run() command is given, the anonymous script is executed. This script can
be run using this command many times without having to re-send it.
Example
run()
Runs the anonymous script.
Also see
script.anonymous (on page 11-157)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals