Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-186 2601B-PULSE-901-01A April 2020
Also see
Adjustment (on page 13-19)
smua.cal.lock() (on page 11-182)
smua.cal.restore() (on page 11-184)
smua.cal.save() (on page 11-184)
smua.cal.unlock() (on page 11-186)
smua.cal.unlock()
This function enables the commands that change calibration settings.
Usage
smua.cal.unlock(password)
Details
This function enables the calibration functions to change the calibration settings.
The password when the instrument is shipped from the factory is "KI0026XX".
Example
smua.cal.unlock("KI0026XX")
Also see
Adjustment (on page 13-19)
smua.cal.lock() (on page 11-182)
smua.cal.password (on page 11-182)
smua.cal.state (on page 11-185)
smua.contact.calibratehi()
This function adjusts the high/sense high contact check measurement.
Usage
smua.contact.calibratehi(cp1Measured, cp1Reference, cp2Measured, cp2Reference)
The value measured by this SMU for point 1
The reference measurement for point 1 as measured externally
The value measured by this SMU for point 2
The reference measurement for point 2 as measured externally