Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-197
Also see
Adjustment (on page 13-19)
smua.cal.lock() (on page 11-182)
smua.cal.restore() (on page 11-184)
smua.cal.save() (on page 11-184)
smua.cal.unlock() (on page 11-186)
smua.source.calibrateY() (on page 11-225)
smua.measure.count
This attribute sets the number of measurements made when a measurement is requested.
Instrument reset
SMU reset
Recall setup
Usage
count = smua.measure.count
smua.measure.count = count
Details
This attribute controls the number of measurements made any time a measurement is requested.
When using a reading buffer with a measure command, this attribute also controls the number of
readings to be stored.
If the count is set to a value greater than 1, any measurement delay set by smua.measure.delay
only occurs before the first measurement, while the smua.measure.interval controls the interval
between successive measurements.
When the pulser is enabled and measurements are not overlapped, this attribute must be less
than 3500.
Example
Sets the measure count to 10.
Also see
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.measure.delay (on page 11-198)
smua.measure.interval (on page 11-202)
smua.measure.overlappedY() (on page 11-205)
smua.measure.Y() (on page 11-209)
smua.reset() (on page 11-222)