Section 10: Instrument programming Model 2601B-PULSE System SourceMeter Instrument Reference Manual
10-46 2601B-PULSE-901-01A April 2020
Master and subordinates
In a TSP-Link
®
system, one of the nodes (instruments) is the master node and the other nodes are
the subordinate nodes. The master node in a TSP-Link
®
system can control the other nodes
(subordinates) in the system.
When any node transitions from local operation to remote operation, it becomes the master of the
system. All other nodes also transition to remote operation and become its subordinates. When any
node transitions from remote operation to local, all other nodes also transition to local operation, and
the master/subordinate relationship between nodes is dissolved.
The expanded system can be stand-alone or computer-based.
Stand-alone system: You can run a script from the front panel of any instrument (node) connected to
the system. When a script is run, all nodes in the system go into remote operation (REM indicators
turn on). The node running the script becomes the master and can control all other nodes, which
become its subordinates. When the script is finished running, all the nodes in the system return to
local operation (REM indicators turn off), and the master/subordinate relationship between nodes
is dissolved.
Computer-based system: You can use a computer and a remote communications interface to any
single node in the system. This node becomes the interface to the entire system. When a command is
sent through this node, all nodes go into remote operation (REM indicators turn on). The node that
receives the command becomes the master and can control all other nodes, which become its
subordinates. In a computer-based system, the master/subordinate relationship between nodes can
only be dissolved by performing an abort operation.
TSP-Link nodes
Each instrument (node) attached to the TSP-Link
®
network must be identified by assigning it a unique
TSP-Link node number.
Commands for remote nodes are stored in the node table. An individual node is accessed as
node[N], where N is the node number assigned to the node.
All TSP-accessible remote commands can be accessed as elements of the specific node. The
following attributes are examples of items you can access:
• node[N].model: The product model number string of the node.
• node[N].revision: The product revision string of the node.
• node[N].serialno: The product serial number string of the node.
You do not need to know the node number of the node that is running a script. The variable
localnode is an alias for the node entry of the node where the script is running. For example, if a
script is running on node 5, you can use the global variable localnode as an alias for node[5]. To
access the product model number for this example, use localnode.model.

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals