EasyManuals Logo

Tektronix Keithley 2601B-PULSE Reference Manual

Tektronix Keithley 2601B-PULSE
849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #430 background imageLoading...
Page #430 background image
Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-72 2601B-PULSE-901-01A April 2020
Usage
func = display.smua.measure.func
display.smua.measure.func = func
func
The type of measurement:
â–ª 0 or display.MEASURE_DCAMPS: Current measurement function
â–ª 1 or display.MEASURE_DCVOLTS: Voltage measurement function
â–ª 2 or display.MEASURE_OHMS: Resistance measurement function
â–ª 3 or display.MEASURE_WATTS: Power measurement function
Details
Selects the measurement function that is displayed on the front panel: Amps, volts, ohms, or watts.
Example
display.smua.measure.func = display.MEASURE_DCAMPS
Selects the current measure function.
Also see
display.smua.limit.func (on page 11-71)
display.trigger.clear()
This function clears the front-panel trigger event detector.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.trigger.clear()
Details
The trigger event detector remembers if an event has been detected since the last
display.trigger.wait() call. This function clears the trigger event detector and discards the
previous history of TRIG key presses.
This attribute also clears the display.trigger.overrun attribute.
Also see
display.trigger.overrun (on page 11-73)
display.trigger.wait() (on page 11-74)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals