smua.measure.delayfactor
This attribute stores a multiplier to the delays that are used when smua.measure.delay is set to
smua.DELAY_AUTO.
Instrument reset
SMU reset
Recall setup
Usage
delayFactor = smua.measure.delayfactor
smua.measure.delayfactor = delayFactor
The delay factor multiplier
Details
The delay factor is only applied when smua.measure.delay = smua.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the
automatic delay.
Setting this attribute to zero disables delays when smua.measure.delay = smua.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay.
Also see
Measure auto delay (on page 4-55)
reset() (on page 11-156)
smua.measure.delay (on page 11-198)
smua.reset() (on page 11-222)
smua.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Instrument reset
SMU reset
Recall setup
Usage
filterCount = smua.measure.filter.count
smua.measure.filter.count = filterCount
The number of readings required for each filtered measurement (1 to 100)
Details
This attribute sets the size of the stack used for filtered measurements.