Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-3
Settings ignored when the pulser is enabled
The following source-measure unit (SMU) settings are ignored when the pulser is enabled:
display.smua.limit.func
smua.measure.autorangeY
smua.measure.autozero
smua.measure.delay
smua.measure.highcrangedelayfactor
smua.measure.lowrangeY
smua.measure.nplc
smua.measure.rangeY
smua.sense
smua.source.autorangeY
smua.source.delay
smua.source.func
smua.source.highc
smua.source.limitY
smua.source.lowrangeY
smua.source.offlimitY
smua.source.rangeY
smua.source.settling
smua.source.sink
smua.trigger.endpulse.action
smua.trigger.endsweep.action
Settings ignored when the pulser is disabled
The following pulser settings do not affect source-measure unit (SMU) operation when the pulser
is disabled:
smua.pulser.measure.aperture
smua.pulser.measure.delay
smua.pulser.protect.sensev
smua.pulser.protect.sourcev
smua.pulser.rangeY