Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-200 2601B-PULSE-901-01A April 2020
Example
smua.measure.filter.count = 10
smua.measure.filter.type = smua.FILTER_MOVING_AVG
smua.measure.filter.enable = smua.FILTER_ON
Sets the filter count to 10.
Sets the filter type to moving average.
Enables the filter.
Also see
Filters (on page 4-77)
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.measure.filter.type (on page 11-201)
smua.reset() (on page 11-222)
smua.measure.filter.enable
This command enables or disables filtered measurements.
Instrument reset
SMU reset
Recall setup
Usage
filterState = smua.measure.filter.enable
smua.measure.filter.enable = filterState
The filter status; set to one of the following values:
â–ª 0 or smua.FILTER_OFF: Disables the filter
â–ª 1 or smua.FILTER_ON: Enables the filter
Details
This command enables or disables the filter.
Example
smua.measure.filter.count = 10
smua.measure.filter.type = smua.FILTER_MOVING_AVG
smua.measure.filter.enable = smua.FILTER_ON
Sets the filter count to 10.
Sets the filter type to moving average.
Enables the filter.
Also see
Filters (on page 4-77)
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.measure.filter.count (on page 11-199)
smua.measure.filter.type (on page 11-201)
smua.reset() (on page 11-222)