Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-207
Also see
Range (on page 4-54)
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.measure.autorangeY (on page 11-194)
smua.pulser.rangeY (on page 11-220)
smua.reset() (on page 11-222)
smua.source.rangeY (on page 11-235)
smua.measure.rel.enableY
This attribute turns relative measurements on or off.
Instrument reset
SMU reset
Recall setup
Usage
relEnable = smua.measure.rel.enableY
smua.measure.rel.enableY = relEnable
Relative measurement control. Set relEnable to one of the following values:
â–ª 0 or smua.REL_OFF: Disables relative measurements
â–ª 1 or smua.REL_ON: Enables relative measurements
SMU measurement function (v = voltage, i = current, r = resistance, p = power)
Details
This attribute enables or disables relative measurements. When relative measurements are enabled,
all subsequent measured readings are offset by the relative offset value specified by
smua.measure.rel.levelY. Each returned measured relative reading is the result of the
following calculation:
Relative reading = Actual measured reading − Relative offset value
Example
smua.measure.rel.enablev = smua.REL_ON
Enables relative voltage measurements.
Also see
Relative offset (on page 4-75)
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.measure.rel.levelY (on page 11-208)
smua.reset() (on page 11-222)