Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-21
Example
bufferCapacity = smua.nvbuffer1.capacity
print(bufferCapacity)
Reads the capacity of dedicated reading
buffer 1 (source-measure unit (SMU)
channel A).
Output:
1.49789e+05
The above output indicates that the buffer
can hold 149789 readings.
Also see
Reading buffers (on page 5-1)
smua.makebuffer() (on page 11-193)
smua.measure.overlappedY() (on page 11-205)
smua.measure.Y() (on page 11-209)
smua.nvbufferY (on page 11-211)
smua.trigger.measure.Y() (on page 11-251)
bufferVar.clear()
This function empties the buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
bufferVar.clear()
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a dedicated
reading buffer
Details
This function clears all readings and related recall attributes from the buffer (for example,
bufferVar.timestamps and bufferVar.statuses) from the specified buffer.
Example
smua.nvbuffer1.clear()
Clears dedicated reading buffer 1
(source-measure unit (SMU) channel A).
Also see
Reading buffers (on page 5-1)
smua.nvbufferY (on page 11-211)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals