Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-174 2601B-PULSE-901-01A April 2020
For example, to set the UTC-5 time zone, you specify the string "5", because UTC-5 is 5 hours
behind UTC and you must add 5 hours to the local time to determine UTC time. To specify the time
zone UTC4, you specify "-4", because UTC4 is 4 hours ahead of UTC and 4 hours must be
subtracted from the local time to determine UTC.
dstStart and dstEnd are strings of the form "MM.w.dw/hh[:mm[:ss]]" that indicate when
daylight savings time begins and ends respectively:
• MM is a number between 1 and 12 that represents the month
• w is a number between 1 and 5 that represents the week in the month
• dw is a number between 0 and 6 that represents the day of the week (where 0 is Sunday)
The rest of the fields represent the time of day that the change takes effect:
• hh represents hours
• mm represents minutes
• ss represents seconds
The minutes and seconds fields are optional.
The week of the month and day of the week fields are not specific dates.
Example
settimezone("8", "1", "3.3.0/02", "11.2.0/02")
settimezone(offset)
Sets offset to equal +8 hours, +1 hour for
DST, starts on Mar 14 at 2:00 am, ends on
Nov 7 at 2:00 am.
Sets local time zone to offset.
Also see
gettimezone() (on page 11-97)
os.time() (on page 11-152)
settime() (on page 11-172)
setup.poweron
This attribute specifies which saved setup to recall when the instrument is turned on.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
0
Usage
id = setup.poweron
setup.poweron = id
id
An integer that specifies the setup to recall when the instrument power is turned on
(0 to 5)
Details
When id = 0, the instrument uses the factory default setup when it is turned on. When id is set to 1
to 5, it uses the setup saved with setup.save().

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals