Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-131
lan.trigger[N].connected
This attribute stores the LAN event connection state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
connected = lan.trigger[N].connected
connected
The LAN event connection state:
â–ª true: Connected
â–ª false: Not connected
N
The LAN event number (1 to 8)
Details
This read-only attribute is set to true when the LAN trigger is connected and ready to send trigger
events following a successful lan.trigger[N].connect() command; if the LAN trigger is not
ready to send trigger events, this value is false.
This attribute is also false when either lan.trigger[N].protocol or
lan.trigger[N].ipaddress attributes are changed or the remote connection closes
the connection.
Example
lan.trigger[1].protocol = lan.MULTICAST
print(lan.trigger[1].connected)
Outputs true if connected, or false if
not connected.
Example output:
false
Also see
lan.trigger[N].connect() (on page 11-130)
lan.trigger[N].ipaddress (on page 11-132)
lan.trigger[N].protocol (on page 11-135)
lan.trigger[N].disconnect()
This function disconnects the LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[N].disconnect()
N
The LAN event number (1 to 8)