Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-191
smua.contact.speed
This attribute stores the speed setting for contact check measurements.
Instrument reset
SMU reset
Recall setup
Usage
speedSetting = smua.contact.speed
smua.contact.speed = speedSetting
The speed setting. Set to one of the following:
â–ª 0 or smua.CONTACT_FAST
â–ª 1 or smua.CONTACT_MEDIUM
â–ª 2 or smua.CONTACT_SLOW
Details
This setting controls the aperture of measurements made for contact check. It does not affect the
smua.measure.nplc aperture setting.
The speed setting can have a dramatic effect on the accuracy of the measurement (see
the specifications).
Contact check is not available when the pulser is enabled. If you need to use the contact check
function in a pulser application, you can disable the pulser, run the contact check measurements,
then enable the pulser.
Example
smua.contact.speed = smua.CONTACT_SLOW
Configure contact check for higher accuracy.
Also see
Contact check connections (on page 4-12)
Contact check measurements (on page 4-12)
reset() (on page 11-156)
smua.contact.check() (on page 11-188)
smua.contact.r() (on page 11-190)
smua.reset() (on page 11-222)