Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-370 2601B-PULSE-901-01A April 2020
tsplink.trigger[N].clear()
This function clears the event detector for a LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
tsplink.trigger[N].clear()
N
The trigger line (1 to 3) to clear
Details
The trigger event detector enters the detected state when an event is detected.
tsplink.trigger[N].clear() clears a trigger event detector, discards the history of the trigger
line, and clears the tsplink.trigger[N].overrun attribute.
Example
tsplink.trigger[2].clear()
Clears trigger event on synchronization line 2.
Also see
tsplink.trigger[N].mode (on page 11-371)
tsplink.trigger[N].overrun (on page 11-372)
tsplink.trigger[N].release() (on page 11-374)
tsplink.trigger[N].stimulus (on page 11-375)
tsplink.trigger[N].wait() (on page 11-376)
tsplink.trigger[N].EVENT_ID
This constant identifies the number that is used for the trigger events.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID = tsplink.trigger[N].EVENT_ID
eventID
The trigger event number
N
The trigger line (1 to 3)
Details
This number is used by the TSP-Link trigger line when it detects an input trigger.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
trigger events from this line.