Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-9
Details
This command enables or disables the beeper. When enabled, a beep signals that a front-panel key
has been pressed. Disabling the beeper also disables front-panel key clicks.
Example
beeper.enable = beeper.ON
beeper.beep(2, 2400)
Enables the beeper and generates a
two-second, 2400 Hz tone.
Also see
beeper.beep() (on page 11-8)
bit.bitand()
This function performs a bitwise logical AND operation on two numbers.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
result = bit.bitand(value1, value2)
result
Result of the logical AND operation
value1
Operand for the logical AND operation
value2
Operand for the logical AND operation
Details
Any fractional parts of value1 and value2 are truncated to form integers. The returned result is
also an integer.
Example
testResult = bit.bitand(10, 9)
print(testResult)
Performs a logical AND operation on decimal 10 (binary 1010)
with decimal 9 (binary 1001), which returns a value of decimal
8 (binary 1000).
Output:
8.00000e+00
Also see
Bit manipulation and logic operations (on page 9-4)
bit.bitor() (on page 11-10)
bit.bitxor() (on page 11-10)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals