Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-168 2601B-PULSE-901-01A April 2020
serial.baud
This attribute configures the baud rate for the RS-232 port.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
9600
Usage
baud = serial.baud
serial.baud = baud
baud
The baud rate (300, 600, 1200, 2400, 4800, 9600, 19200, 38400, 57600, or 115200)
Details
A new baud rate setting takes effect when the command to change it is processed.
Allow ample time for the command to be processed before attempting to communicate with the
instrument again. If possible, set the baud rate from one of the other command interfaces or from the
front panel.
The reset function has no effect on data bits.
Example
serial.baud = 1200
Sets the baud rate to 1200.
Also see
RS-232 interface operation (on page 2-32)
serial.databits (on page 11-169)
serial.flowcontrol (on page 11-170)
serial.parity (on page 11-170)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals