Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 11: TSP command reference
2601B-PULSE-901-01A April 2020 11-393
userstring.get()
This function retrieves a user-defined string from nonvolatile memory.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
value = userstring.get("name")
value
The value of the user-defined string key-value pair
name
The name (key) of the user-defined string
Details
This function retrieves the string that is associated with name from nonvolatile memory.
Example
userstring.add("assetnumber", "236")
value = userstring.get("assetnumber")
print(value)
Create the user-defined string assetnumber, set to a value of 236.
Read the value associated with the user-defined string named assetnumber.
Store it in a variable called value, then print the variable value.
Output:
236
Also see
userstring.add() (on page 11-390)
userstring.catalog() (on page 11-391)
userstring.delete() (on page 11-392)
waitcomplete()
This function waits for all previously started overlapped commands to complete.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
waitcomplete()
waitcomplete(group)
group
Specifies which TSP-Link group on which to wait

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals