Section 11: TSP command reference Model 2601B-PULSE System SourceMeter Instrument Reference Manual
11-374 2601B-PULSE-901-01A April 2020
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Usage
tsplink.trigger[N].release()
The trigger line (1 to 3)
Details
Releases a trigger that was asserted with an indefinite pulse width. It also releases a trigger that was
latched in response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Also see
tsplink.trigger[N].assert() (on page 11-369)
tsplink.trigger[N].clear() (on page 11-370)
tsplink.trigger[N].mode (on page 11-371)
tsplink.trigger[N].overrun (on page 11-372)
tsplink.trigger[N].pulsewidth (on page 11-373)
tsplink.trigger[N].stimulus (on page 11-375)
tsplink.trigger[N].wait() (on page 11-376)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger attributes to their factory defaults.
Usage
tsplink.trigger[N].reset()
The trigger line (1 to 3)
Details
The tsplink.trigger[N].reset() function resets the following attributes to their factory
defaults:
â–ª tsplink.trigger[N].mode
â–ª tsplink.trigger[N].stimulus
â–ª tsplink.trigger[N].pulsewidth
This also clears tsplink.trigger[N].overrun.