smua.measure.highcrangedelayfactor
This attribute contains a delay multiplier that is only used during range changes when the high-capacitance mode
is active.
Instrument reset
SMU reset
Recall setup
Usage
delayFactor = smua.measure.highcrangedelayfactor
smua.measure.highcrangedelayfactor = delayFactor
The delay factor; set to a value between 1 and 99
Details
This delay multiplier is only active when the high-capacitance mode is active.
Example
smua.measure.highcrangedelayfactor = 5
Increases the delay used during range changes by a
factor of 5.
Also see
High-capacitance mode (on page 4-80)
reset() (on page 11-156)
setup.recall() (on page 11-175)
smua.reset() (on page 11-222)
smua.source.highc (on page 11-228)
smua.measure.interval
This attribute sets the interval between multiple measurements.
Instrument reset
SMU reset
Recall setup
Usage
interval = smua.measure.interval
smua.measure.interval = interval
The interval value (in seconds); set to a value between 0 and 1