EasyManuals Logo

Tektronix Keithley 2601B-PULSE Reference Manual

Tektronix Keithley 2601B-PULSE
849 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #790 background imageLoading...
Page #790 background image
Section 13: Calibration Model 2601B-PULSE System SourceMeter Instrument Reference Manual
13-32 2601B-PULSE-901-01A April 2020
14. Turn on the output:
smua.source.output = smua.OUTPUT_ON
15. Allow the readings to settle, then get both the multimeter and 2601B-PULSE current readings at the
negative zero value (the 2601B-PULSE measurement is not necessary if this step is being done on the
CALA sense). The two measurements should be made as close as possible in time. Use this command
for the 2601B-PULSE:
Z_rdg = smua.measure.i()
16. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
17. Set the source output to the negative full-scale value, for example:
smua.source.leveli = -90e-3
18. Turn on the output:
smua.source.output = smua.OUTPUT_ON
19. Allow the readings to settle, then get both the multimeter and 2601B-PULSE current readings at the
negative full-scale output value (the 2601B-PULSE measurement is not necessary if this step is being
done on the CALA sense mode). The two measurements should be made as close as possible in time.
Use this command for the 2601B-PULSE:
FS_rdg = smua.measure.i()
20. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
21. Send the source calibration command using the -range, -zero and -FS multimeter readings, and -zero
and -FS source values for the parameters:
smua.source.calibratei(-range, src_Z, DMM_Z_rdg, src_FS, DMM_FS_rdg)
Where:
-range
= The negative of the present adjustment range
src_Z
= The zero 2601B-PULSE source output value
DMM_Z_rdg
= The zero DMM measurement
src_FS
= The FS 2601B-PULSE source output value
DMM_FS_rdg
= The FS DMM measurement
Typical values for the 100 mA range:
smua.source.calibratei(-100e-3, -1e-30, -1e-6, -90e-3, -89.2e-3)
22. If this step is not on the CALA sense mode, send the measure calibration command using the
multimeter and 2601B-PULSE readings, and range setting for the parameters:
smua.measure.calibratei(-range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg)
Where:
-range
= The negative of the present adjustment range
Z_rdg
= The zero 2601B-PULSE measurement
DMM_Z_rdg
= The zero DMM measurement
FS_rdg
= The FS 2601B-PULSE measurement
DMM_FS_rdg
= The FS DMM measurement
Typical 100 mA range values:
smua.measure.calibratei(-100e-3, -1e-5, -1e-6, -91e-3, -89.2e-3)

Table of Contents

Other manuals for Tektronix Keithley 2601B-PULSE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 2601B-PULSE and is the answer not in the manual?

Tektronix Keithley 2601B-PULSE Specifications

General IconGeneral
BrandTektronix
ModelKeithley 2601B-PULSE
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals