Model 2601B-PULSE System SourceMeter Instrument Reference Manual Section 13: Calibration
2601B-PULSE-901-01A April 2020 13-45
40. Send the current measure calibration command using the range, averaged zero and averaged full scale
digitizer readings, and averaged zero and averaged full scale source values for the parameters:
smua.pulser.measure.calibratei(range, DMM_Z_Rdg, digi_Z_rdg, FS_IRdg,
digi_FS_rdg)
Where:
range
= The present adjustment range
DMM_Z_Rdg
= The -zero scale 2601B-PULSE measurement value
digi_Z_rdg
= The -zero scale digitizer measurement
FS_IRdg
= The -full scale 2601B-PULSE measurement value
digi_FS_rdg
= The -full scale digitizer measurement
Typical values for the -1 A measurements range:
smua.pulser.measure.calibratei(-1, -0.01, -0.011, -0.75, -0.76)
41. Send the voltage measure calibration command using the range, averaged zero and averaged full scale
digitizer readings, and averaged zero and averaged full scale source values for the parameters:
smua.pulser.source.calibratei(range, ZS_VRdg, digi_Z_rdg, FS_VRdg, digi_FS_rdg)
Where:
range
= The present adjustment range
ZS_VRdg
= The -zero scale 2601B-PULSE measurement value
digi_Z_rdg
= The -zero scale digitizer measurement
FS_VRdg
= The -full scale 2601B-PULSE measurement value
digi_FS_rdg
= The -full scale digitizer measurement
Typical values for the -1 A measurements range:
smua.pulser.measure.calibratei(-1, -0.01, -0.011, -0.75, -0.76)
E. Select automatic polarity mode:
smua.cal.polarity = smua.CAL_AUTO