Remote Control Commands
R&S
®
SMBV100B
841User Manual 1178.4460.02 ─ 03
[:SOURce<hw>]:LFOutput:FREQuency:MANual <Manual>
Sets the frequency of the subsequent sweep step if LFO:SWE:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any value within the setting range, where:
STARt is set with [:SOURce<hw>]:LFOutput:FREQuency:
STARt
STOP is set with [:SOURce<hw>]:LFOutput:FREQuency:
STOP
Range:  STARt to STOP
Increment:  0.1
*RST:  1000
Example:  See Example "Setup an LF sweep" on page 837.
Manual operation:  See " Current Frequency" on page 328
[:SOURce<hw>]:LFOutput:FREQuency:MODE <Mode>
Sets the mode for the output of the LF generator frequency, and determines the com-
mands to be used for frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Sets the fixed-frequency mode. CW and FIXed are synonyms.
To set the output frequency, use command [:SOURce]:
LFOutput<ch>:FREQuency
SWEep
Sets sweep mode.
To set the frequency, use the commands:
[:SOURce<hw>]:LFOutput:FREQuency:STARt and [:
SOURce<hw>]:LFOutput:FREQuency:STOP
Or [:SOURce<hw>]:LFOutput:FREQuency:MANual
*RST: 
CW
Example:  See Example "Setup an LF sweep" on page 837.
Manual operation:  See "State (LF frequency sweep)" on page 327
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
[:SOURce<hw>]:LFOutput:FREQuency:STARt <Start>
Sets the start/stop frequency for [:SOURce<hw>]:LFOutput:FREQuency:MODE
 SWEep.
SOURce Subsystem