CALCulate Subsystem 3
N1911A/1912A P-Series Power Meters Programming Guide 169
CALCulate[1]|2|3|4:GAIN[:MAGNitude] <numeric_value>
This command is used to enter a value for the display offset on the
specified window/measurement. The display offset is applied to the
measurement signal after any math calculation.
Entering a value using this command automatically turns the
CALCulate[1]|2|3|4:GAIN:STATe command to ON.
Syntax
Parameters
Item Description/Default Range of Values
numeric_value A numeric value for the display offset:
• DEF: the default value is 0 dB
• MIN: –100.000 dB
• MAX: +100.000 dB
–100.000 to +100.000 dB
DEF
MIN
MAX
CALC
:MAGN
:GAIN
Space
?
numeric_value
DEF
MIN
MAX
MIN
MAX
Space
1
2
3
4