EasyManua.ls Logo

Agilent Technologies N1911A

Agilent Technologies N1911A
803 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 353
PSTatistic[1]|2:CCDF:TRACe:POWer:PEAK?
This command is used to retrieve peak power value of Channel A or
Channel B trace.
Syntax
Example
Error Messages
If no power sensor is connected, error –241 "Hardware missing" occurs.
If sensor/sensors connected are not P- Series sensors, error –241
"Hardware missing" occurs.
If measurement speed setting is FAST, error –221 "Settings conflict"
occurs.
If command executed in other window besides CCDF window, error
–221 "Settings conflict: Requires CCDF window" occurs.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
:POW
PST
2
1
:CCDF :TRAC
?
:PEAK
PST:CCDF:TRAC:POW:PEAK? This command returns the peak power
value of Channel A trace.

Table of Contents

Other manuals for Agilent Technologies N1911A

Related product manuals