9 PSTatistic Subsystem
344 N1911A/1912A P-Series Power Meters Programming Guide
PSTatistic[1]|2:CCDF:TABle?
This command is used to return the data in CCDF table, average input
power, probability at the average input power, peak to average power ratio
and sample count.
NOTE
This command will return 10 scalar results in the following order:
1 Average input power (in dBm)
2 Probability at the average input power (in %)
3 Power level (power to average power ratio) that has 10 % of the power (in dB)
4 Power level (power to average power ratio) that has 1 % of the power (in dB)
5 Power level (power to average power ratio) that has 0.1 % of the power (in dB)
6 Power level (power to average power ratio) that has 0.01 % of the power (in dB)
7 Power level (power to average power ratio) that has 0.001 % of the power (in dB)
8 Power level (power to average power ratio) that has 0.0001 % of the power (in dB)
9 Peak to average power ratio (in dB)
10 Sample count
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
• Free run acquisition mode is selected
• NORMal or DOUBle measurement speed setting is chosen

Table of Contents

Other manuals for Agilent Technologies N1911A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N1911A and is the answer not in the manual?

Agilent Technologies N1911A Specifications

General IconGeneral
BrandAgilent Technologies
ModelN1911A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals