PSTatistic Subsystem 9
N1911A/1912A P-Series Power Meters Programming Guide 315
PSTatistic:CCDF:MARKer[1]|2:DATa?
This command is used to retrieve the power and probability values at the
current marker position at any trace the marker is positioned.
Syntax
Example
Reset Condition
On reset, the marker 1 and marker 2 will be set back to their default
positions.
NOTE
This command is only applicable when P-Series sensors are present and the following
conditions are met:
Free-run acquisition mode is selected
NORMal or DOUBle measurement speed setting is chosen
PST
:CCDF
2
:DAT
1
:MARK
?
PST:CCDF:MARK1:DAT? This command returns the power and
probability values at marker 1.

Table of Contents