11 STATus Subsystem
480 N1911A/1912A P-Series Power Meters Programming Guide
STATus:OPERation:SENSe[:SUMMary]
The operation status sense summary register set contains information on
the status of the power sensors.
The following bits in these registers are used by the power meter:
These bits are set when the power meter is reading data from the
E- Series power sensor or N8480 Series power sensor EEPROM.
These bits are cleared when the power meter is not reading data from the
E- Series power sensor or N8480 Series power sensor EEPROM.
Syntax
Bit Number Decimal
Weight
Definition
0 - Not used
1 2 Channel A SENSe Status
2 4 Channel B SENSe Status (N1912A only)
3-15 - Not used
STAT
:OPER
:SENS
:SUMM